Product Details
 

 
   
AquaSensors Dissolved Ozone DataStick

 

 

Product Features

 

:: Clark cell technology

:: Pre-calibrated (no field calibration required)

:: Plug & play sensor heads

:: Rugged, foul resistant membrane

:: Electrode protection options

:: Simple membrane cap replacement

:: Direct data reporting (24-bit)

:: Plug & play industrial communications adapters

 

Specifications

 

Product Specification

Description

Measurement

Range: 0 ppm – 10 ppm

System

Resolution: 0.01 ppm

Performance*

Accuracy: 2%

Step Response Time

90% in < 90 seconds

Operational Environment

Temperature Range: –5 °C to 50 °C
Maximum Pressure: 65 psig @ 50 °C
Maximum Flow Rate: 10 ft/second

Power
Requirements #

Voltage Range: 10 to 30 VDC
Maximum Power: 200 mW
Typical Power: 120 mW

Construction

Process Electrodes: Gold cathode, silver
anode, silver reference (3 electrode polarographic clark cell)
Guard: Platinum Membrane: PFA Teflon®
O-rings: Viton® (other materials available)
Sensor Head Material: PEEK
DataStick Material: 316 stainless steel, PEEK or CPVC
Weight:
1.2 lbs (PEEK or CPVC)
2.6 lbs (316 stainless steel)

Units of Measure

Measurement Units: ppm
Temperature Units: °C, °F

Calibration**

Sample: 1 point
Zero: 1 point
Temperature: 1 point

Compensation
Options

Temperature:
Automatic from –5 °C to 50 °C

Other Configuration
Options

Sensor Filter: 0 to 100 seconds
Temperature Filter: 0 to 100 seconds

Approvals and
Ratings

Immunity & Emissions: CE Certified 89/336/EEC: CISPER 11,EN61000 (-4-2,-4-3,-4-4,-4-6, 4-8)
Safety: cULus Listed; 367G E303570
Hazardous Locations: Haz Loc Class 1,
Division 2, Groups A, B, C, D.
Max Ambient 50 °C

* Typical at 25 °C Performance unaffected by cable length
# Class II DC power supply required
** Ozone and Temperature are precalibrated at the factory

 

Applications

Drinking water treatment ,Wastewater treatment, Food and beverage sanitization, Bottled water production, Packaged water systems, Pulp & paper bleaching, Pharmaceutical Cooling water, Semiconductor wash water

 

Ordering Information (Conductivity Sensor)

 

Part No.

Description

DS-b-t

DataStick

Body Material (b)

1 = 316 Stainless Steel
2 = CPVC
3 = PEEK

Mounting (t)

1 = 1 inch NPT front/back
2 = Flow Chamber Snap-ring

CA-b-nw-x-y

Communications Adapter

Body Material (b)

1 = 316 Stainless Steel
2 = CPVC
3 = PEEK

Communications (nw)

1A = RS232 ASCII
2B = Modbus® RTU
2A = Modbus RS232
4B = CANopen
7R = Ethernet
5R = DeviceNet
6R = Profibus DP
8R = USB

Cable Length (x)

1 = 10 feet
2 = 20 feet
3 = 30 feet

Cable Termination (y)

A = Stripped Wires

OZ-b-t-x

Ozone Sensor Head

Body Material (b)

3 = PEEK

Electrode Type (t)

1 = Gold/Silver

Sensor Tip (x)

B = Process Flat
C = Face Seal for Flow Chamber

Note: Consult for additional configurations and accessories

 

Thermo Scientific AV38 DataStick Local Monitor and Controller

 

Features & Benefits

 

:: Connects to any DataStick system

:: Convenient ¼ DIN enclosure

:: Display sensor and temperature data

:: Menu functions: Calibration, configuration and diagnostics

:: Digital protocol options for remote measurement, calibration, configuration and diagnostics

:: Two 4 to 20 mA current outputs

:: Two Form C relays for alarm, control and wash

:: PID controller for use with pH and conductivity

:: 7 conduit knockouts for easy cablin

 

Specifications

 

Product Specification

Description

Mounting Options

Wall-mount Enclosure: NEMA 4X
Panel-mount: ¼ or ½ DIN NEMA
4X panel mount available
Display Only: with NEMA 4X gasket

User Interface

Display: 2 line by 16 character LCD
Backlight: High contrast green
Keypad: 4 directional arrows, MENU,
ESCAPE and ENTER

Environmental Conditions

Ambient Operating Temperature Range:
–20 to 65 °C
Maximum Relative Humidity:
95% non-condensing

DataStick Sensor Connections

Supports 1 to 255 Modbus® RTU DataStick Sensors of any type
Supported Sensors: pH, ORP, two electrode conductivity, two electrode resistivity, toroidal
conductivity, dissolved oxygen (ppm and ppb) dissolved ozone, drinking water and wide-range turbidity, suspended solids and free chlorine

4 to 20 Milliamp Current Output Options

Assignable parameters
PID function when pH or conductivity parameter selected Loop power provided by AV38

Relays

Two 1 amp Form C relays with NO and
NC contacts
Assignable parameters
Configured for alarm, control and timer
functions

Digital
Communications

Options to Host Computer Network:
MODBUS® RTU, DeviceNet,
Modbus RS232, RS232 ASCII, CANopen,
Ethernet IP, Modbus TCP, TCP/IP
Others: Consult factory

Power Supply

Input: 10 to 30 VDC or 100 to 240 VAC
(optional)
Power: 2W at 25°C

Approvals

Meets CE requirements for heavy
industrial use

Note: Consult for additional configurations and accessories

 

Order Information

 

Part No.

Description

AV38-v-w-x-y-z

Local Monitor/Controller

Current Outputs (v)

B = One
C = Two

Mounting (w)

B = ¼ DIN NEMA 4X Wall-mount Enclosure

Communications (x)

0 = None
1 = RS232 ASCII
2 = Modbus RS232
4 = MODBUS RTU (for Modbus + Profibus)
5 = DeviceNet
6 = CANopen
7 = Ethernet IP, Modbus TCP, TCP/IP

Relays (y)

A = None
C = Two

Power (z)

1 = 24 VDC
2 = 100 to 240 VAC

 

Download Brochure

Dissolved Ozone System DataStick
AV38 Datastick Local Monitor And Controller

     
© Copyright 2010 ITLIPL Designed by DBS   Home | About Us | Products | What's New | Career | Clients | Contact Us